Mass Resolution
The mass resolution is the dimensionless ratio M/DM.
Low mass resolution refers to M/DM= 300, enough to separate ions that differ by at least one amu.
Sometimes, however, High Mass Resolution is needed to separate ions that have
the same nominal mass, like 30Si+H (30.981597 amu) and 31P (30.973763 amu) for
instance.
At Probion, we run magnetic sector Ion Mass Spectrometers. These allow M/DM
values as high as 10 000.
(Almost) all common interferences can be solved easily :
The mass resolution is the dimensionless ratio M/DM
|
Ion
|
Interference
|
MATRIX
|
M/DM
|
|
27Al+
|
11B16O+
|
Si
|
1190
|
|
27Al+
|
12C21H3+
|
Si
|
645
|
|
28Si-
|
27Al1H-
|
AlGaAs
|
2250
|
|
31P-
|
30Si1H-
|
Si
|
3955
|
|
32S-
|
31P1H-
|
InP
|
3360
|
|
39K+
|
11B28Si+
|
Si, BPSG
|
1730
|
|
56Fe+
|
28Si2+
|
Si
|
2955
|
|
75As-/+
|
74Ge1H-/+
|
Ge, SiGe
|
10115
|
|
|
Probion Analysis
37, rue de Fontenay - 92220
Bagneux - France
Tel : 33 (0)1 46 12 05 00 - Fax : 33 (0)1 46 54 10 33
info
probion.fr
|
| ©Probion Analysis 2006, All Rights Reserved. |